Electrical Troubleshooting - Teledyne VISION G2 CLASSIC 6 User Manual

Table of Contents

Advertisement

Electrical Troubleshooting

Problem
Display is dark
USB functions not working
System does not make
noise
Screen is illuminated but
system not responsive
Vision G2 Testers User Guide—74-108-800 Rev. H—12 June 2019
Possible Causes
Tester is not turned on.
Tester is not plugged in.
USB Drive is not formatted.
USB drive is damaged.
Monument board is
damaged.
Speaker volume set to 0.
Display board is damaged.
System has frozen.
Touchscreen requires
calibration.
Troubleshooting
Recommended Action
Use power switch to turn
tester on.
Ensure tester is plugged into
an outlet that is delivering
power.
The Teledyne Hanson USB
drive has already been
formatted. If using another
USB drive, format using a PC.
Use another USB drive.
Replace monument board.
Use Display screen to raise
speaker volume.
Replace display board.
Note last action, reboot
system. Report issue to THR
Technical Support.
If possible, recalibrate through
display screen. If display
screen cannot be reached,
reboot system and touch the
screen during the logo splash
screen.
125 of 158