Small Volume Vessels

1.
Confirm the small volume vessels and mini paddles
are clean and in good condition. Install them in their
designated positions.
2.
On the tester, touch the temperature to bring up the
set temperature screen. Enter the desired
temperature (+ 0.10 °C of the desired vessel
temperature is recommended) and touch OK.
3.
Fill vessels with media. Pull the spindles up to the
highest level. Lower drive head if using an Elite 8.
4.
Place covers on vessel.
5.
Once the media has stabilized, ensure the protocol
is loaded by checking the upper left corner of the
screen. If the correct protocol is loaded, do nothing.
If it is not, touch the upper left corner of the screen,
then touch Exit and select the protocol to be used
for testing.
6.
Touch the Start button in the lower right corner of
the screen. Enter the header information as
appropriate, then touch Next (
7.
Lift the cover and drop the dosage into the first
position.
8.
Touch Run.
9.
Once the appropriate offset has passed, drop the
next dosage into position 2, and so on.
Recommended offset time between positions is 2
minutes.
Dissolution Testing
).