Spindle Shaft Height Adjustment; Oving And Torage - Teledyne VISION G2 CLASSIC 6 User Manual

Table of Contents

Advertisement

Vision G2 Testers User Guide—74-108-800 Rev. H—12 June 2019

Spindle Shaft Height Adjustment

1.
Install either the paddle or basket shaft with basket
into the spindle shaft.
2.
Place a height set gauge (from the Height Set and
Check kit) in the bottom of the vessel of the affected
position.
3.
Gently lower the paddle or basket so it rests on top
of the height gauge.
4.
Using a hex key, loosen the bolt on the spindle shaft
clamp.
5.
Adjust the spindle clamp so that it is properly
engaged to the spindle. Make sure that the paddle
or basket is still resting on top of the height gauge.
6.
Tighten the spindle clamp using the same hex key.
7.
Confirm the height of any apparatus used in that
spindle position. If not all apparatuses meet the
necessary height requirements, contact THR
Technical Support.
Moving and Storage
If the Vision tester will not be used for a short period of time (3
days or more), the instrument should be cleaned and the power
should be turned off. If the tester will not be used for an
extended period of time (30 days or more), the instrument
should be cleaned, drained, covered, and unplugged from the
power source.
The instrument is heavy and great care should be taken when
moving it. If the instrument is moved to a new location,
revalidation should be considered.
Two people are required to lift the Classic 6 and/or the Elite 8
dissolution tester. Lifting should be done with one person
situated on each side of the instrument. Lift the equipment by
the base frame.
Warning: Lifting without assistance may cause injury.
Maintenance
123 of 158

Advertisement

Table of Contents
loading

This manual is also suitable for:

Vision g2 elite 8

Table of Contents