Manual (Or 3Rd Party Autosampling) Using Automag With Digital Temperature Probes (Dtps) - Teledyne VISION G2 CLASSIC 6 User Manual

Table of Contents

Advertisement

Vision G2 Testers User Guide—74-108-800 Rev. H—12 June 2019
Manual (or 3rd Party Autosampling) Using AutoMag
with Digital Temperature Probes (DTPs)
1.
Confirm the vessels, paddles, or baskets and basket
shafts are clean and in good condition. Install them
in their designated positions.
2.
Confirm the heights of apparatus from the vessel
bottom using a height gauge.
3.
On the tester, touch the temperature to bring up the
set temperature screen. Enter the desired
temperature (+ 0.10 °C of the desired vessel
temperature is recommended) and touch OK.
4.
Raise the apparatus out of the vessels by raising the
drive head.
5.
Fill the vessels with media.
6.
Prepare the apparatus:
a. Baskets - Elite 8
i. Lower the spindles so they are at their
ii. Cover the vessels
iii. Allow the media to stabilize.
b. Paddles
i. Lower the drive head to the testing
ii. Cover the vessels and allow the media
c. Paddles with ADD Cover
i. Lower the drive head to the testing
ii. Close the chamber of the ADD cover
iii. Cover the vessels and allow the media
Dissolution Testing
lowest point; do not move the drive
head.
position.
to stabilize.
position.
by pressing the hinged portion on the
bottom of the cover closed.
to stabilize.