Mechanical Troubleshooting
Problem
Excessive vibration in
system
Paddle/Basket heights not
correct
Vision G2 Testers User Guide—74-108-800 Rev. H—12 June 2019
Possible Causes
Belts are too tight or too
loose.
Tester is not level.
Alignment nut of high stand
tester may be loose.
Improper operation
Clamp has loosened or
spindle shaft replaced.
Troubleshooting
Recommended Action
Replace or adjust belts
according to maintenance
procedure in this manual.
Ensure all feet of tester are
secure on bench.
Tighten alignment nut.
Confirm that spindle clamp
and spindle are properly
engaged.
Confirm that paddle or basket
shaft is properly tightened in
spindle shaft.
Perform the spindle height
adjustment procedure listed in
the maintenance section.