Appendix B: Networking

Appendix B: Networking

Networks

A network is a series of devices that are connected together and able to communicate
with each other. The Vision G2 dissolution testers and samplers communicate with
each other as a network using TCP/IP protocols.
In order for the Vision G2 AutoPlus to communicate with the Vision dissolution testers,
it must be connected through a network switch. The AutoPlus and testers each
communicate through a TCP/IP protocol which requires a unique IP address, default
gateway address, and subnet mask.
A network PostScript capable printer may be connected directly to the switch if the
operator wants to use this printer while keeping the system isolated from the larger
corporate network.

Isolated Networks

If the operator wishes to have an isolated system, then the switch does not need to be
connected to anything else, other than the AutoPlus and testers.
Switch
Tester 
Tester 
Sampler
1
2
An isolated THR Vision G2 system should be set up as follows:
Vision G2 AutoPlus
IP Address: 192.168.1.10
Default Gateway: 192.168.1.1
Subnet Mask: 255.255.255.0
Vision Dissolution Tester 1
IP Address: 192.168.1.20
Default Gateway: 192.168.1.1
Subnet Mask: 255.255.255.0
Vision G2 Testers User Guide—74-108-800 Rev. H—12 June 2019
141 of 158