Appendix C: Waters Alliance Diss. System
4.
Touch the Magazine Travel field and set the value to what is desired for
the test.
5.
Exit the protocol to save it.
6.
Start a test using the 2690D/2695D or Empower, and the tester will move
the AutoMag (or SuperMag) appropriately when collecting a sample.

Limitations

The system has the following limitations when running a test using the 2690D/2695D
and Transfer Module.
The event table will be ignored in the WTM protocol.
The sampling events, temperature, infinity test speed, and auto print interval range
are all restricted by Empower.
The maximum label length is 20 characters instead of 25.
The speed tolerance is either off or USP ± 4% instead of programmable from 1 to
10%.
The temperature tolerance is either off or USP ± 0.5 °C instead of programmable
from 0.1 °C to 5.00 °C.
The IP address of the second Vision G2 tester must be sequential with the first
tester.
If a Waters Transfer Module protocol is not present, the protocol ID "Protocol" will be
used when a test is running.
Waters Empower will not record the vessel temperatures. A separate printer for the
tester is required.
If ADD covers are used, the tester will delay the start of the test by 3 seconds plus
the Auto Dosage Delivery Delay in the protocol. For example, if the Auto
Dosage Delivery Delay was set to 5 seconds, the tester would start the test 8
seconds after the HPLC. Users can correct this by adjusting their sample points
accordingly.
Vision G2 Testers User Guide—74-108-800 Rev. H—12 June 2019