AutoMag/SuperMag Troubleshooting

Problem
Magazine is not moving
the proper distance
Magazine is not moving at
all.
Vessel temperature
probes not working.
When measuring
temperature, Probe Error
or --.-- is displayed
Vision G2 Testers User Guide—74-108-800 Rev. H—12 June 2019
Possible Causes
Magazine may not be
calibrated.
Protocol not set up for proper
sample distance.
Magazine not configured.
Component failure.
Vessel temperature probes
not configured.
Vessel temperature probes
not connected.
Component failure.
Probe is disconnected.
Probe is damaged.
Troubleshooting
Recommended Action
Perform magazine calibration.
Change Magazine Travel in
protocol.
Ensure protocol makes use of
magazine.
Ensure magazine is set up in
Device Setup.
Contact THR Technical
Support.
Ensure probes are configured
for 6, 7 or 8 probes as
appropriate in device setup.
Ensure that probes are
connected to tester main
board.
Contact THR Technical
Support.
Reconnect temperature probe.
Contact THR Technical
Support.