Installation; Location - Teledyne VISION G2 CLASSIC 6 User Manual

Table of Contents

Advertisement

Installation

Installation

Location

This section provides information on determining if a location is
suited to a Vision G2 dissolution tester. Consideration should
be made as to the final system layout design prior to placing any
equipment on the bench.
Environmental Requirements
The location should be clean and free of any issues which may
influence dissolution testing. Issues which may affect
dissolution results include vibration, significant heat, or
significant cold.
Space Requirements
See the specifications section to determine space requirement
based on the size of the equipment.
The bench should be level and flat.
The bench should be capable of supporting the full weight
(including water) of each machine as laid out in the
specifications section.
The first Vision G2 dissolution instrument should be placed to
the left of an autosampler, if used. Additional testers should go
to the right of the autosampler.
Electrical Requirements
Each Vision G2 dissolution tester requires two grounded outlets
within 1 meter (3 ft.) of the location of the system; one for the
dissolution unit and one for the heater.
The Vision G2 testers are available in one configuration, 100 V
through 240 V. The voltage is automatically set when the
voltage is supplied to the external power supply.
Power cords should meet the following specifications: IEC 320-
C13 connector; SJT 18AWG3C, 2 x 0.824 mm2, 60 °C, 300 V.
Note that USA-style power cords are provided with the
equipment. It is the responsibility of the end user to provide
power cords that meet that country's electrical connection
requirements. The power cords also meet or exceed the above
specifications.
Vision G2 Testers User Guide—74-108-800 Rev. H—12 June 2019
24 of 158