Microbial Filter 654 St - Dräger Julian Instructions For Use Manual

Anaesthetic workstation
Hide thumbs Also See for Julian:
Table of Contents

Advertisement

Microbial filter 654 St

Wipe the surface with a disposable cloth.
Do not clean with solvents such as naphtha, alcohol or
ether.
Do not clean the filter in cleaning and disinfecting
machines.
Do not immerse the filter in disinfectant solution.
Sterilize the filter at max. 134 °C in a hot steam
sterilizer.
Do not use ethylene oxide sterilization!
Do not sterilize when connected!
Mark each sterilization in the corresponding field on
the filter housing (e.g. with a water-resistant pen).
The filter can be sterilized up to 24 times.
The filter must be discarded:
– if the pressure rise in the resistance test is greater
than 2 mbar.
– after being sterilized 24 times at the most.
The filter must be sterilized once more in hot steam at
134 °C after use and can then be disposed of as
plastic with ordinary domestic waste.
Note the hygiene regulations of the hospital.
All the components are listed in the care list for Juian on
pages 104 and 105, together with the corresponding
intervals and appropriate means.
Disinfecting / cleaning / sterilizing
Care
103