Test Iso Connector For Vaporizers; Leak Test; Test Vaporizer On Associated Anaesthetic Machine - Dräger Vapor 2000 Technical Documentation Manual

Table of Contents

Advertisement

7

Test ISO connector for vaporizers

• Check that the male cone at the vaporizer inlet and the female cone at
the vaporizer outlet are connected.
8

Leak test

8.1

Test vaporizer on associated anaesthetic machine

• Lock the lock catch or cap.
• Switch on the vaporizer (> 0.2 vol.%).
• Depressurise the AIR system and the N
• Close all flow control valves.
• Connect the mixed gas socket to the pressure gauge.
• Set the O
flow control valve such that a pressure of 30 mbar is
2
indicated on the pressure gauge.
• Check that the gas flow is < 0.12 L/min.
This test list does not replace the inspections and servicing to be carried out by Dräger Medical AG & Co. KGaA
5327.400 Vapor 2000
O system.
2
04.99
Test list
Page 30