3.8.2
Designing a customized test protocol
The f ollowing sections describe how to setup your own test protocol in the HIT440 setup. All setup options
f or all tests will be described even though you most likely will only need a f ew. This section is theref ore to be
seen as a work of ref erence where single f unctions can be looked up rather than a step-by-step guide where
all steps need to be explored.
1) Enter the REM440 setup by pressing Menu/Setup/HIT440 setup.
2) Press New.
3) A pop-up will ask whether the new test protocol should be based on a copy of the currently selected
test protocol. When asked by the message box:
Yes:
The system will create a copy of the current test and allow you to add new tests, make
changes, and save it under a new name.
No:
The Selected tests section will be blank, and you can create a new test f rom scratch
.
4) Name the test protocol setup in the Selected protocol section.
D-0004576-Q – 2022/05
Affinity/Equinox2.0 – Additional Information
2