Technical Specifications Of The Ac440 Software - Interacoustics Affinity 2.0 Additional Information

Hide thumbs Also See for Affinity 2.0:
Table of Contents

Advertisement

1.19 Technical specifications of the AC440 software

Medical CE-mark:
Audiometer standards:
Transducers & calibration:
Air conduction
DD45
TDH39
HDA300
DD450
HDA280
E.A.R Tone 5A
IP30
Bone conduction
B71
B81
Free field
High frequency
Effective masking
Patient response switch:
Patient communication:
Monitor:
Stimuli:
Tone
Warble tone
Wave file
Masking
Narrow band noise:
White noise:
Speech Noise.
Presentation
Intensity
Steps
Accuracy
Extended range
f unction
Frequency
Distortion (THD)
D-0004576-Q – 2022/05
Affinity/Equinox2.0 – Additional Information
The CE-mark in combination with MD symbol indicates that
Interacoustics A/S meets the requirements of the Medical Device
Regulation (EU) 2017/745 Annex I
Approval of the quality system is made by TÜV – identif ication no. 0123.
Tone: IEC60645-1/ANSI S3.6 Type 1
Speech: IEC60645-2/ANSI S3.6 Type A or A-E
Calibration inf ormation and instructions are in the Service manual.
Check the accompanying Appendix f or RETSPL levels f or transducers
ISO 389-1 2017, ANSI S3.6-2018
ISO 389-1 2017, ANSI S3.6-2018
PTB report 1.61.4066893/13
ISO 389-8 2004, ANSI S3.6-2018
PTB report 2004
ISO 389-2 1998, ANSI S3.6-2018
ISO 389-2 1998, ANSI S3.6-2018
Placement: Mastoid
ISO 389-3 2016, ANSI S3.6-2018
ISO 389-3 2016, ANSI S3.6-2018
ISO 389-7 2005, ANSI S3.6-2010
ISO 389-5 2004, ANSI S3.6-2010
ISO 389-4 1994, ANSI S3.6-2010
Handheld push button.
Talk Forward and Talk Back.
Output through external earphone or speaker.
Pure tone, Warble tone, NB, SN, WN, TEN noise
125-20000Hz separated in two ranges 125-8000Hz and 8000-20000Hz.
Resolution 1/2-1/24 octave.
1-10 Hz sine +/- 5% modulation
44100Hz sampling, 16 bits, 2 channels
Automatic selection of narrow band noise (or white noise) f or tone
presentation and speech noise f or speech presentation.
IEC 60645-1:2001, 5/12 Octave f ilter with the same centre f requency
resolution as pure Tone.
80-20000Hz measured with constant bandwidth
IEC 60645-2:1993 125-6000Hz f alling 12dB/octave above 1KHz +/-5dB
Manual or Reverse. Single or multiple pulses. pulse time adjustable f rom
200mS-5000mS in 50mS steps. Simultaneous or alternating.
Check the accompanying Appendix f or maximum output levels
Available Intensity Steps is 1, 2 or 5dB
Sound pressure levels: ± 2 dB
Vibration f orce levels: ± 5 dB
If not activated, the Air Conduction output will be limited to 20 dB below
maximum output.
Range: 125Hz to 8kHz (Optional High Frequency: 8 kHz to 20 kHz)
Accuracy: Better than ± 1 %
Sound pressure levels: below 1.5 %
Vibration f orce levels: below 3 %.
Headband Static Force 4.5N ±0.5N
Headband Static Force 4.5N ±0.5N
Headband Static Force 8,8N ±0.5N
Headband Static Force 10N ±0.5N
Headband Static Force 5N ±0.5N
Headband Static Force 5.4N ±0.5N
Headband Static Force 5.4N ±0.5N
Page 143