3.7.6 Attack/recovery time

The Attack time is the time it takes f or the hearing aid compressor to react to an increase in input signal. As
the output will gradually move towards the f inal gain value it has been decided to def ine the attack time as
the time it takes the hearing aid to stabilize the output within 2 or 3 dB according to the IEC 6018-2 and ANSI
S3.22 respectively. The attack time is commonly 5 ms but can be longer.
The Recovery is the time taken f or the hearing aid compressor to react to a decrease in input level. As the
output will gradually move towards the f inal gain value it has been decided to def ine the recovery time as the
time it takes the hearing aid to stabilize the output within 2 (f ollowing an input decrease f rom 80 to 55 dB) or
4 dB (f ollowing an input decrease f rom 90 to 55 dB) according to the IEC 118-2 and ANSI S3.22
respectively. The release time is commonly 20 ms but can be longer.
Measuring the attack and recovery time gives you an idea of the time constants within the hearing aid and
how f ast it reacts to incoming sounds. If the recovery time is too short the gain will vary too much during
speech that shif ts quickly in amplitude, phase, and f requency. This will distort the signal. In hearing aids with
short attack time and long recovery time the distortion will be minor. The main disadvantage in this case will
be extremely brief sounds such as a door slamming will cause the gain to decrease as the attack time is
short. However, the gain will stay low f or a while due to the long recovery time which may result in the client
missing inf ormation during this period.
Note: If using the IEC standard this test should be perf ormed with the volume control in its maximum position
(e.g., bef ore Ref erence Test Gain)