REIG measurements

REIG (Real-Ear Insertion Gain): The REIG accounts f or the gain across f requencies provided by the
hearing aid alone. This is obtained by subtracting the REUR f rom the REAR or the REUG f rom the REAG.
The REIG is most of ten used to verif y if a given hearing aid setting is providing an acceptable level of gain
compared to the prescribed REIG target.
Test procedure:
11
2
12
6
9
14
1) Open the REM440 module by selecting the REM tab.
2) If needed select a test protocol in the List of Protocols. If you do not select a protocol the system
will automatically select a standard protocol.
3) Explain the procedure to the client.
4) Perf orm an otoscopic examination.
5) Mount the probe tube on the in-situ headset and place it on the client who should be seated ½-1
meters f rom the speaker.
Ensure that the probe tubes are calibrated bef ore placing them in the ear canal (see section 0 about
tube calibration).
6) Measure the REUG/REUR (see section 0) or select a predicted REUG/REUR (see section 0).
Note: Since the REIG is a calculation the system will require that you perf orm the REUR bef ore
perf orming the REIG.
D-0004576-Q – 2022/05
Affinity/Equinox2.0 – Additional Information
REAR – REUR = REIG
13