Binaural Rem Measurements - Interacoustics Affinity 2.0 Additional Information

Hide thumbs Also See for Affinity 2.0:
Table of Contents

Advertisement

10) Ensure that the correct Ear is selected
11) Choose the pref erred Stimulus in the corresponding dropdown list .
If you have calibrated f or open f it only the signals used f or this calibration are available in the
stimulus selection dropdown (see section 0).
12) Select input level using the Input level slider
13) Click START and f ollow the instructions on screen.
14) Place the client in f ront of the loudspeaker as shown on the picture. Click OK and f ront microphone
curve will be recorded. New instructions will immediately appear.
15) Turn the client around and click OK. The directionality curve will then be f inalised as shown above
Note: The continuous measurement button is disabled f or Directionality. Measurement time must theref ore
be set in the REM440 setup (see section 2.6.2.7).
2.6.8

Binaural REM measurements

This is a f eature which allows binaural measurement of real ear measurements simultaneously. This makes
it easier and quicker to perf orm this measurement.
This test involves the placement of the probe tube microphone in each of the patient's ears at the same time
and ref erencing is perf ormed with only one of the ref erence microphones. As a consequence, this test
requires that the patient is orientated at 0° f rom the speaker to ensure that there is equal distance to the
speaker f rom each ear.
Binaural REM can only be used with stimuli which do not require active ref erencing during the measurement.
Theref ore, Binaural REM is not suitable f or measurement with Warble Tone, Pink Noise, Random Noise,
Pure tone, and Narrowband Noise.
No specif ic protocol is required to perf orm Binaural REM as it uses the settings f rom your existing protocols.
However, a protocol can be def aulted to start up in Binaural REM measuring mode by enabling the below
setting in the REM protocol.
38
2.0
If using the Affinity
with external speaker the client does not need to be turned around (see section 2.6.2.7 for directionality setup
options)
D-0004576-Q – 2022/05
Affinity/Equinox2.0 – Additional Information