®
R&S
NGE100B
Example:
Usage:
[SOURce:]POWer:PROTection[:STATe] <arg0>
This command defines or queries the OPP state for the previous selected channel.
Parameters:
<arg0>
Example:
[SOURce:]POWer:PROTection:CLEar
This command resets the OPP state of the selected channel. If the OPP has tripped
the red blinking OPP message on the display will be cleared for the selected channel.
Usage:
6.5.10 Measurement Commands
MEASure[:SCALar]:CURRent[:DC]?
This command queries the measured current value of the previous selected channel.
Return values:
<Current>
Example:
Usage:
MEASure[:SCALar]:POWer?
This command queries the measured power value of the previous selected channel.
User Manual 5601.1343.02 ─ 04
INST OUT1
POW:PROT:TRIP? -> 0
OPP state of Ch 1 has not tripped.
Query only
ON | OFF | 1 | 0
ON | 1
OPP is activated.
OFF | 0
OPP is deactivated.
INST OUT1
POW:PROT ON
OPP of Ch 1 is activated.
POW:PROT?
Return the OPP state of the selected channel.
Event
<numeric_value>
Return the measured current.
Default unit: A
MEAS:CURR? -> 1.000
Query only
Remote Control
Remote Control - Commands
88