3.3 Sample-by-sample Measurement (6-cell manual mode)

3.3.1 Determining Concentration of Solution

This instrument is used for generating a calibration curve and determining the concentration
of an unknown sample, and for entering a calibration factor and determining the
concentration.
GUIDE: When auto start is specified, the system automatically sets up the conditions and
advances to the measurement screen phase of sequence after the power switch is
turned on. For how to specify auto start, refer to section 4.5..
1.
Startup of This Product
Start this product. (For the starting procedure, refer to
section 2.3.1..)
2.
Setting of Measuring Conditions
3.
Setting of Sample Conditions
4.
Setting of Calibration Curve Conditions
5.
Setting of Calibration Curve Data
* For steps 2 to 5, refer to section 3.2.1..
6.
Setting of 6-cell Conditions
"6 Cell Mode" is not indicated if OFF is specified
NOTE:
for Mode on the 6-cell mode setting screen for
use of the single cell holder or rectangular long
path cell holder. (For details, refer to section
2.5.4..) When the auto sipper is connected,
"Sipper" is indicated in place of "6 Cell Mode".
(1) To set the 6-cell conditions, press the <5> key (6 Cell)
and then [ENTER] key. Or select "6 Cell" with the [▲]
or [▼] key and press the [ENTER] key or [►] key.
3 - 98
*