Hitachi U-5100 Instruction Manual page 135

Ratio beam spectrophotometer
Table of Contents

Advertisement

6.
Saving of measuring conditions
GUIDE: When the set measuring conditions need not be
saved, proceed to step 7.
(1) To save the set measuring conditions, press the <5>
key (Save Method) and then [ENTER] key. Or select
"Save Method" with the [▲] or [▼] key and press the
[ENTER] key or [►] key.
(2) The measuring condition saving (Save Method) screen
(Fig. 3-42) appears.
Meas.PARAM.
1 CheckB (at 500nm)
2 Quality Check
3 DNA meas.
File name Measurement Saving
File Name:_
File Name [Max: 20 characters]
Fig. 3-42
(3) When other measuring conditions are saved in
advance, a list of saved measuring conditions appears
on the main menu. At this time, the file name,
measurement mode (PHT: Photometry, WLS:
Wavelength Scan, TMS: Time Scan) and saving date
are indicated.
(4) For the measuring conditions to be saved, enter a file
name within 20 characters and press the [ENTER] key.
3 - 71
PHT 2010/10/10
PHT 2010/03/30
PHT 2010/04/29
mode
date
( 1/ 1)
Measuring Condition Saving
(Save Method) Screen
3.2.3
CELL A