(3)
After start of measurement, you must wait for about 80
minutes till the end of all measurements. Upon completion
of measurement, the screen after completion of automatic
performance check (Fig. 5-30) is displayed. OK appears
when the result of each check meets the judgment standard
or NG appears if it does not meet the standard. If NG is
indicated, take a corrective measure, referring to section
5.1.1 to 5.1.6.
Performance
(Auto)
WL Accuracy (484.6 nm)
WL Repeatability
Noise Level (RMS)
Baseline Flatness
Baseline Stability
Hardware
Execute it? :_
□ 1 Yes □ 2 No
Fig. 5-30 Screen after Automatic Performance Check
5 - 23
CELL A
-0.2 nm
OK
±0.10 nm
OK
0.0001 ABS
OK
±0.000 ABS
OK
0.0002 ABS
OK
OK