Hitachi U-5100 Instruction Manual page 319

Ratio beam spectrophotometer
Table of Contents

Advertisement

Error Message
Xe Lamp Error
Abnormality is found
when turning on the
lamp.
Error occurred
Wavelength
when WL was
initialization is
initialized
abnormal.
WL Motor Error
Wavelength
initialization is
abnormal.
WL CALIB. Error
Abnormality is found in
wavelength calibration
with the bright line of
lamp.
WL Correction
Abnormality is found in
Error
wavelength correction.
Data Save Error
RAM does not work
normally.
6 Cell Initialize
Initialization of 6-cell
Error
holder position is
abnormal.
6 Cell Move Error
Abnormality is found in
detection of cell
position.
6 - 11
Cause
Turn on the lamp again.
If no improvement can be
seen, contact the
maintenance service
company authorized by
Hitachi High-
Technologies.
Retry wavelength
initialization. If no
improvement can be
seen, contact the
maintenance service
company authorized by
Hitachi High-
Technologies.
Retry wavelength
initialization. If no
improvement can be
seen, contact the
maintenance service
company authorized by
Hitachi High-
Technologies.
Retry wavelength
calibration. If no
improvement can be
seen, contact the
maintenance service
company authorized by
Hitachi High-
Technologies.
Retry wavelength
correction. If no
improvement can be
seen, contact the
maintenance service
company authorized by
Hitachi High-
Technologies.
Retry data saving. If no
improvement can be
seen, contact the
maintenance service
company authorized by
Hitachi High-
Technologies.
Mount this holder
properly and retry
initialization. Or if any
holder other than the 6-
cell holder is used, set 6-
cell condition, referring to
section 2.5.4..
Check the sample
compartment to see if
there is anything that
hinders rotation of 6 cells.
6.8
(cont'd)
Remedy