3.3 Measurement

3.3.2 Concentration Measurement

Given below is the general measurement procedure for the
quantitative calculation to create calibration curves.
(1) Switching to measurement screen
After setting the conditions, select the [STD Measurement]
button on the screen to proceed to the measurement screen.
Fig. 3-6 Screen to Proceed to Measurement Screen
NOTE: The settings of "Sipping Time", "Delay Time", and "Air
Purge" are common to the standard measurement and the
sample measurement. Pay attention to the viscosity and
temperature of each solution in setting.
(2) Sipping of autozero sample
After moving to the measurement screen, the autozero
execution window is displayed. Place the sipping nozzle in the
sample solution for autozero, push the sipping lever to sip the
solution, to measure the autozero automatically.
To execute an autozero between an interval of the
measurements for standard solutions or sample solutions, press
[autozero] icon and follow the guidance to execute the
autozero.
NOTE: Pressing the [OK] button on the screen without pushing the
sipping lever permits executing the autozero without
sipping.