3.2 Automatic Continuous Measurement (6-cell auto mode)
NOTE:
• If a file of the same name as entered already exists in
the same measurement mode, the following guidance
appears.
Already exists. Overwrite it? :__
□ 1 Yes □ 2 No
For overwriting, press the <1> key (Yes). To avoid
overwriting, press the <2> key (No) and rename the file
to be saved.
• A total of 50 method files can be saved for all
measurement modes.
• If 50 files are already registered, the following message
appears.
No.of files is full.
In this case, delete files referring to section 4.2.2 and
retry saving.
• The method for which auto start (for details, refer to
section 4.5) is set is preceded by "*".
(5) Return to the Photometry screen by pressing the
[RETURN] key.
10. Measurement of Standard Solutions
GUIDE: When 2) Factor is selected for Curve Type in
step 4, standard solution measurement is not
carried out. So proceed to step 11.
(1) To determine the measuring conditions set in the
preceding steps, press the <9> key (Meas. Screen)
and then [ENTER] key. Or press the [MEAS.
SCREEN] key.
(2) The standard setting screen (Fig. 3-13) appears.
3 - 26