3.4 Measurement with Enlarged Display Screen (monitor)
3.
Presentation of Enlarged Display Screen (monitor)
The enlarged display screen (Fig. 3-103) appears. At this
time, the lamp turns on automatically and measurement
starts. In the upper part of the monitor, the current
wavelength setting is displayed, and a photometric value in
the lower part. This photometric value is updated at
intervals of 2.5 sec. To display the absorbance, press the
<1> key (ABS) and the <2> key (%T) to display the
transmittance, then press the [ENTER] key.
When changing the wavelength to be used for
measurement, press the [GO TO WL] key, enter a new
wavelength and press the [ENTER] key.
Data Display
WL:
1100.0
Data:
Select function. : _
1)ABS 2)%T 3)Lamp ON
Fig. 3-103 Enlarged Display Screen
The enlarged display screen is provided with the automatic
lamp off function, which serves to prevent the lamp staying
on unnecessarily for a long time. After the lapse of
automatic lamp off time, the lamp is automatically turned off
and "-.---" is indicated for the photometric value as shown in
Fig. 3-104. To display the photometric value again, press
the <3> key (Lamp ON) and then [ENTER] key.
For the auto lamp off time (Auto Lamp OFF), "5 min" is set
as default. For changing this setting, refer to section 2.5.3.
3 - 148
nm
0.001
ABS
CELL 1