Hitachi U-5100 Instruction Manual page 180

Ratio beam spectrophotometer
Table of Contents

Advertisement

3.3 Sample-by-sample Measurement (6-cell manual mode)
5.
Setting of Printing Conditions
6.
Saving of Measuring Conditions
* For steps 5 and 6, refer to section 3.2.2..
Upon completion of setting, press the <6> key (Meas. Screen)
and then [ENTER] key or press the [MEAS. SCREEN] key to
determine the measuring conditions.
7.
Measurement of Sample Solutions
(1) The auto zero execution screen (Fig. 3-74) appears.
Set an auto zero sample to cell A. Auto zero is
executed by pressing the [AUTOZERO] key.
Photometry
[Info]
Set autozero sample to Cell A.
AUTOZERO:Execute Autozero; CLEAR:Cancel
Fig. 3-74 Auto Zero Execution Screen
(2) The sample measurement screen (Fig. 3-75) appears.
Set the sample indicated in the guidance area in the
sample compartment. Pressing the [START] key
starts measurement at the current cell position.
In 6-cell manual mode, the photometric value at
NOTE:
the cell position indicated on the screen is
measured. Before measurement, therefore,
check if the cell position where the sample has
been set coincides with the cell position
indicated on the screen.
3 - 116
*
600.0 nm 0.000 ABS
CELL A

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents