4.6.4 How to Return 6-Cell Holder
1.
Removal of Present Cell Holder
Firstly, remove the mask for micro cell when mounted to the
single cell holder. Then, remove the single cell holder or
rectangular long-path cell holder, whichever mounted, by
loosening the screw shown in the figure below.
Fig. 4-77 Removal of Single Cell Holder
Fig. 4-78 Removal of Rectangular Long-Path Cell Holder
2.
Mounting of 6-cell Holder
Mount the 6-cell holder, referring to the 6-cell holder
mounting method described in "b." of 2.4.5.
3.
Change of 6-cell Mode Setting
The 6-cell mode setting needs to be changed.
Turn the 6-cell mode setting from OFF to ON, referring to
section 2.5.4.
4 - 59
4.6.4
Screw