Photometry
1 Meas.PARAM.
2 Sample
3 6 Cell
4 Print
5 Save Method
6 Meas.Screen
Select function. : _
□ 1 Auto □ 2 Manual
Fig. 3-73 6-cell Condition (6 Cell) Screen
(3) To set each item on the 6-cell condition (6 Cell)
screen, press the [ENTER] key. Or select each item
with the [▲] or [▼] key and make your setting
according to the guidance. For details of each
parameter, refer to Table 3-46.
After selecting "Manual" for 6 Cell Mode, return to the
Photometry screen by pressing the [RETURN] key or [◄]
key .
Table 3-46 "6 Cell" Setting Parameter
Setting Item
6 Cell Mode
Specify a 6-cell operation mode for measurement.
[1] Auto
In this mode, a sample is set at each position
of the 6-cell holder beforehand and the holder
is automatically turned to automatically
perform from auto zero to standard/sample
measurement. It is recommended to use this
mode if you have 6 cells and there are many
samples to be measured. (For details of the
Auto mode, refer to section 3.2.2..)
[2] Manual
In this mode, standard/sample measurement
is carried out by using one position (cell 1 to 5
selectable) of the 6-cell holder. To eliminate
the influence of instrument drift (baseline
fluctuation) during measurement, auto zero
can easily be executed by setting a solution
for auto zero to cell A. In normal usage, set
a standard or sample solution to cell 1.
3 - 115
CELL A
6 Cell Mode
Auto
SAMP.Autozero
ON
Autozero int.
5
No.of Sample
1
Description
3.3.2