5.1.2 Wavelength Repeatability
(1)
Display the screen for each item of instrument performance
check (Fig. 5-7). For display method, follow the procedure
starting from section 5.1.. Using the [▲]/[▼] key, move
the cursor to Repeatability of WL, and then select [1] Yes.
Wavelength repeatability will be checked. If check is not
desired actually, select [2] No.
Performance
WL Accuracy (484.6 nm)
WL Accuracy (229.0 nm)
WL Accuracy (822.8 nm)
WL Repeatability
Noise Level (RMS)
Baseline Flatness
Baseline Stability
Hardware
Print Report
Execute(Manual)? ::_
□ 1 Yes □ 2 No
Fig. 5-7 Screen for Each Item of Instrument
Performance Check
(2)
The guidance for informing the parameter setting process
(Fig. 5-8) appears, and parameters will be set and
wavelength repeatability will be checked.
Setting PARAM....
Fig. 5-8 Guidance for Parameter Setting Process
(3)
After the check of wavelength repeatability, the screen after
this check is displayed as shown in Fig. 5-9.
Measurement result is indicated on the wavelength
repeatability line followed by OK when the judgment
standard is met or NG if not met. If NG is indicated, check
the wavelength repeatability again in the correct procedure.
If NG recurs, refer to section 6.8 .
5 - 7
5.1.2
CELL A