Hitachi U-5100 Instruction Manual page 176

Ratio beam spectrophotometer
Table of Contents

Advertisement

3.3 Sample-by-sample Measurement (6-cell manual mode)
Setting Item
[2] Curve
You can generate a calibration curve with
the measured standards and confirm it.
When this function is selected, the
following screen appears.
Photometry/Curve
R = 0.9999, R2 = 0.9999
0.893
ABS
0.001
Select function. : _
□ 1 Scale □ 2 Factor □ 3 STD
For details of the function, refer to section
4.3.
[1] Scale :
[2] Factor :
[3] STD
[3] Save
You can save the measured data.
When this function is selected, the
following screen appears.
Save Data
1 NO3-(Hitachi Taro)
2 Cr6+(Hitachi Hanako)
Select function. : _
□ 1 Save □ 2 Sort
[1] Save :
[2] Sort :
3 - 112
Description
+
+
+
+
+
+
+
+
+
0.000
CONC
90.000
Allows you to change the
scale of calibration curve.
Indicates the factor,
correlation coefficient and
determination coefficient of
calibration curve.
:
Indicates the result of
standard solution
measurement. Also allows
remeasurement of a
standard solution, deletion
of standard data
unnecessary for generation
of a calibration curve or
retrieval of deleted data.
PHT 2010/10/10
PHT 2010/03/30
◄ ► ( 1/ 1)
Allows you to save data under
a new file name.
Allows you to rearrange the
currently saved data in the
order of file names or dates.
(cont'd)
CELL A
CELL A