3.2 Automatic Continuous Measurement (6-cell auto mode)
Table 3-29 "6 Cell" Setting Parameters
Setting Item
6 Cell Mode
Specify a 6-cell operation mode for measurement.
[1] Auto
In this mode, a sample is set at each position of the
6-cell holder beforehand and the holder is
automatically turned to automatically perform from
auto zero to standard/sample measurement. It is
recommended to use this mode if you have 6 cells
and there are many samples to be measured.
[2] Manual
In this mode, standard/sample measurement is
carried out by using one position (cell 1 to 5
selectable) of the 6-cell holder. To eliminate the
influence of instrument drift (baseline fluctuation)
during measurement, auto zero can easily be
executed by setting a solution for auto zero to cell A.
In normal usage, set a standard or sample solution to
cell 1.
(For details of the Manual mode, refer to section
3.3.3..)
SAMP.
You can select an auto zero sample in sample
Autozero
measurement or specify avoidance of automatic
execution of auto zero.
[1] ON
Auto zero is automatically carried out in sample
measurement. The auto zero interval is set in
Autozero int.
[2] OFF
Auto zero is not carried out automatically in sample
measurement. You can carry out auto zero
manually.
Autozero int.
[1] 5
Auto zero is automatically carried out for every 5
measurements.
[2] 1
Auto zero is automatically carried out for each
sample.
No. of
Set the number of samples to be measured.
Sample
This parameter is settable in a range of 1 to 150.
3 - 68