Hitachi U-5100 Instruction Manual page 77

Ratio beam spectrophotometer
Table of Contents

Advertisement

Photometry
1 Meas.PARAM.
2 Sample
3 Curve
4 Curve Data
5 System
6 Print
7 Save Method
8 6 Cell
9 Meas.Screen
Input value. : _
STD1 [0.000 - 9999]
Fig. 3-6 Calibration Curve Data (Curve Data)
Screen (Curve Type: 1st Order)
(3) Enter the concentration of each standard (standard
solution). After that, press the [ENTER] key.
You can enter the concentration of each standard by
using the [▲] or [▼] key. Here, key in the
concentration values according to the number of
standards entered on the calibration curve condition
(Curve) screen. The concentration can be entered in
a range of 0.000 to 9999.
[When Curve - Curve Type: Factor is selected]
(2)' The calibration curve data (Curve Data) screen (Curve
Type: Factor) (Fig. 3-7) appears.
Photometry
1 Meas.PARAM.
2 Sample
3 Curve
4 Curve Data
5 System
6 Print
7 Save Method
8 6 Cell
9 Meas.Screen
Input value. : _
A0 [±99999 – 0.0000]
Fig. 3-7 Calibration Curve Data (Curve Data)
Screen (Curve Type: Factor)
(3)' Enter calibration curve factors. After that, press the
[ENTER] key. You can enter each factor by using the
[▲] or [▼] key. The factor can be entered in 5
significant digits in a range of 0.0000 to ±99999.
For how to set the calibration curve factors, refer to
Table 3-4.
3 - 13
CELL A
STD
CONC
1
0.000
2
10.00
3
20.00
CELL A
A0
0.0000
A1
1.0000
3.2.1