5.1 Check with Instrument Main Unit Alone
5. PERFORMANCE CHECK
This section pertains to the performance check method for
checking whether the instrument provides the specified
performance. For performance check with a pen type low-
pressure mercury lamp, a separately available option is required.
For mounting method of this lamp, refer to the instruction manual
for the separately available option.
(1)
After startup of the instrument, press the [MAIN MENU] key
to display the Main Menu screen (Fig. 5-1). Confirm that
nothing is contained in the sample compartment and close
the lid. For selection of performance check, press the <7>
key (Performance) and then the [ENTER] key or press the
[▲]/[▼] key to select Performance followed by the [ENTER]
key.
Main Menu
1 Photometry
2 WL Scan
3 Time Scan
4 Method Menu
5 Data Menu
6 System Setup
7 Performance
8 Data Display
Select function. : _
(2)
The instrument performance check screen (Fig. 5-2)
appears.
Performance
1 Performance Check(Auto)
2 Performance Check(Manual)
3 Performance Check(Option)
Select function. : _
Fig. 5-2 Instrument Performance Check Screen
5 - 1
Fig. 5-1 Main Menu Screen
5.1
CELL A
CELL A