Calibration Data - wtw Oxi 3310 IDS Operating Manual

Digital meter for digital ids-d.o. sensors
Table of Contents

Advertisement

Oxi 3310 IDS (Oxi 3315)
Displaying the cali-
bration record
Calibration evalua-
tion
ba75944e02
10/2014
6.
Using <> <>, set the correction multiplier to adjust the displayed
concentration value to the nominal value (value of the comparison mea-
surement). Subsequently, accept the correction multiplier with
<ENTER>.
The meter switches to the measured value display.
The status display [Factor] is active.

5.3.6 Calibration data

The calibration record is automatically transmitted to the interface
after calibrating.
The calibration record of the last calibration is to be found under the menu item,
Calibration / Calibration record. To open it in the measured value display, press
the <CAL_> key.
The calibration records of the last 10 calibrations are to be found in the menu,
Calibration / Calibration data storage / Display. To open the Calibration menu,
press the <ENTER> key in the measured value display.
Menu item
Calibration /
Calibration data stor-
age / Display
Calibration /
Calibration data stor-
age / Output to
RS232/USB
After calibration, the meter automatically evaluates the current status of the cal-
Setting/
Explanation
function
-
Displays the calibration records.
Further options:
 Scroll through the calibration
records with <><>.
 Output the displayed calibra-
tion record to the interface with
<F2>/[USB-Ausgabe].
 Output all calibration records to
the interface with
<F2__>[USB-Ausgabe].
 Quit the display with <F1>/
[Zurück] or <ENTER>.
 Switch directly to the mea-
sured value display with <M>.
-
Outputs the calibration data mem-
ory to the interface
Dissolved oxygen
31