Calibration Data - wtw VWR MU 6100 H Operating Manual

Table of Contents

Advertisement

pH value
Displays the calibra-
tion data
Displaying the cali-
bration data memory
Calibration evalua-
tion
36
6.2.6

Calibration data

The calibration record is automatically transmitted to the interface
after calibrating.
The calibration record of the last calibration is to be found under the menu item,
<OK>
/ Calibration / Calibration record. To open it, press the
the measured value display.
Subsequently, you can transmit the displayed calibration data to the interface,
e.g. to a PC, with the
The calibration records of the last calibrations are to be found in the menu,
<OK>
/ Calibration / Calibration data storage.
Menu item
Calibration / Calibra-
tion data storage /
Display
Calibration / Calibra-
tion data storage /
Output to USB
After calibrating, the meter automatically evaluates the calibration. The zero
point and slope are evaluated separately. The worse evaluation of both is taken
into account. The evaluation appears on the display and in the calibration
<F2>
/[USB output] key.
Setting/
function
-
-
<CAL__>
Explanation
Displays the calibration record.
Further options:
 Scroll through the calibration
<
><
records with
 Output the displayed calibra-
tion record to the interface with
<F2>
/[USB output].
 Quit the display with
<OK>
[Back] or
.
 Switch directly to the mea-
sured value display with
<MODE>
.
Outputs the calibration records to
the interface.