Oxi 3315
Displaying the cali-
bration record
Calibration evalua-
tion
ba75944e02
03/2012
6.
Using <> <>, set the correction factor to adjust the displayed con-
centration value to the nominal value (value of the comparison mea-
surement). Subsequently, accept the correction factor with <ENTER>.
The measuring instrument switches to the measured value display.
The status display [Factor] is active.

5.3.6 Calibration data

The calibration record is automatically transmitted to the interface
after calibrating.
The calibration record of the last calibration is to be found under the menu item,
Calibration / Calibration record. To open it in the measured value display, press
the <CAL_> key.
The calibration records of the last 10 calibrations are to be found in the menu,
Calibration / Calibration data storage / Display. To open the Calibration menu,
press the <ENTER> key in the measured value display.
Menu item
Calibration /
Calibration data stor-
age /Display
Calibration /
Calibration data stor-
age /
Output to RS232/
USB
After the calibration, the meter automatically evaluates the current status of the
Setting/
Description
function
-
Displays the calibration records.
Further options:
 Scroll through the calibration
records with <><>.
 Output the displayed calibra-
tion record to the interface with
<F2>/[USB output].
 Output all calibration records to
the interface with <F2__>[USB
output].
 Quit the display with <F1>/
[Zurück] or <ENTER>.
 Switch directly to the mea-
sured value display with <M>.
-
Outputs the calibration data stor-
age to the interface
Dissolved oxygen
31