100 | DX-D 300 | Problem solving
DR Detector is Exceeding the Maximum Working
Temperature
Details
Cause
Brief Solution

DR detector must be recalibrated

Details
Cause
Brief Solu-
tion

DR Detector Problem

Details
Cause
Brief Solution
0172G EN 20230627 1651
A message is displayed on NX indicating that the DR De-
tector is exceeding the maximum working temperature.
Due to ambient temperature conditions and the number
of acquired images, the DR Detector's internal tempera-
ture may become too high.
1. Power off the DR Detector.
2. Leave the DR Detector unpowered for at least one
hour.
3. Stop the NX workstation.
4. Power on the DR Detector.
5. Start the NX workstation.
A message is displayed indicating that the DR detector must be
recalibrated.
A DR detector must be recalibrated at regular interval.
Follow the instructions in the user manual to calibrate the DR
detector:
• DR Detector Calibration Key User Manual, document 0134
An error message is displayed on NX indicating a prob-
lem related to the DR detector.
-
1. Power off the DR detector.
2. Stop the NX workstation.
3. Power on the DR detector.
4. Start the NX workstation.