The locks to secure the DR detectors cannot be
released
Details
The locks to secure the DR detectors are not released when start-
ing DR 100s or when pressing the action button on the software
console. The DR detector cannot be removed from the storage bin.
Cause
Mechanical or software failure.
Brief
Solu-
tion
Figure 68: Manually releasing the locks to secure the DR detectors
1. Remove the plastic cap (1).
2. Insert a screw M4x35 (2) where the plastic cap was attached.
3. Pull the screw while releasing the lock.
4. Remove the DR detector from the storage bin.
5. Reactivate the lock.
6. Try to release the lock by pressing the action button on the
7. Remove the screw and attach the plastic cap.
2
1
software console. If the lock is not released, contact your local
service organization.
DR 100s | Problem solving | 157
0411D EN 20220627 1040