Prepare the Full Leg Full Spine configuration

In the examination room, position the X-Ray system, the Full Leg Full Spine
wall stand:
1. Insert the DR detector it in the Full Leg Full Spine wall stand.
The DR Detector must be inserted in portrait position.
2. Adjust the height of the DR detector holder relative to the position of the
patient and the bodypart that is examined.
3. Position the X-ray tube for exposing the Full Leg Full Spine wall stand.
Rotate the X-ray tube 90°.
4. Set the source image distance large enough for the X-ray field to cover at
least one partial image, as indicated on the front panel of the DR detector
holder.
Figure 10: Configuration
5. Adjust the height aligning the center of the X-ray beam to the center of the
DR detector holder.
6. Apply collimation using the collimator on the X-ray tube.
Adjust the width according to the region of interest. Adjust the height to
make sure the exposed area on the DR Detector holder has a height of at
least 45 cm.
7. Move the DR detector to the first position.
Depending on the region of interest, this can be the top most or middle
position in the DR detector holder.
Press the button of the control unit to move the DR detector to the next
position.
During movement, the leds of the target position on the control unit is
blinking. Once the position is reached, the led stays on and the DR
detector can be moved to the next position by pressing the button again.
Full Leg Full Spine DR Retrofit System | Basic Workflow | 41