6.5.1 Setting of stack lengths

Enter the lengths of the stack (inner diameter) where the receiver unit and the transmitter unit are attached. It
does not include the lengths of companion flange. This value has a direct effect on the measurement value, so
be sure to enter correctly. Otherwise, measurement value will not be properly displayed or output. Enter the
value in the millimeter. (Input range is from 500 to 15000mm.)
Explanation of concentration output
The analyzer is calculated and output as a standard of the 1m of stack lengths (optical path lengths).
Based on Lambert-Beer Law, absorption concentration varies depending on the stack lengths, so be sure to
enter the stack lengths correctly.
Receiver unit
ampanion flange
Parameter Setting
Path lengths
Average Period
Reset Avg.Gas
Key Lock
Disp.Off
Date/Time
2006-04-01
12:00:00
Parameter Setting
Display/Output
Input Signal
AO Select
Password
*****
Stack lengths (mm)
2007-4-1
12:00:00
01000
mm
Min.
01
HC ℓ
OFF
ON
09
Min.
07 / 01 / 01
00:00:00
- 48 -
key.
NT
Parameter Setting
Path lengths
Average Period
Reset Avg.Gas
Key Lock
NT
Disp.Off
Date/Time
Point the
to "Path Lengths" by the
and the
key, and press the
2007- 4 -1
12:00:00
01000
mm
Min.
01
HC ℓ
OFF
ON
09
Min.
07 / 01 / 01
00:00:00
Transmitter unit
key
key.
INZ-TN1ZSS-E

Advertisement

Table of Contents
loading

Table of Contents