4.12 Adjustment of fine adjustment screw

This screw is used for adjustment when the light path length in the stack is long and when the quantity of
light misses too much at the time of retightening in the procedure in "4.11 Adjustment of light quantity".
When each retightening is finished, perform fine adjustment by pressing with this screw.
When you use the fine adjustment screw, set it to the B flange before mounting the receiver and the trans-
mitter boxes on the angle adjustment unit. Beware that the fine adjustment screw cannot be set after mount-
ing these boxes.

4.13.1 AC power connection

AC power connecting terminal is positioned at the lower left of the control unit (see Fig. 4 - 4.). Use the AC
cable for which flame resistance is 600V, 1.25mm
φ 10mm or less.
When connecting the AC cable to the input terminal, the ground cable should be longer than the L, N line.
The cable length should be adjusted so that when the AC cable is pulled from the external, the L, N line will
be removed from the terminal block first, and the ground cable will be removed last.
POWER
M4 screw terminal
C cable
Flame resistance: 600V, 1.25mm
Cable out diameter size: 10mm or less.
(standard accessory)

4.13 Connecting to control unit

L
N
PE
C
V
SOURCE
The ground cable should be longer than the N, L line.
(16 WG) or more
2
2
(16AWG) or more, and the cable out diameter size is
C power terminal
Fig. 4-15
- 22 -
INZ-TN1ZSS-E