Input switching of the remote range (DI4 terminal)
Apply rectangular waveform (pulse width: 2.0 sec or more, pulse interval: 2.0 sec or more) to the remote
range input terminal (DI4 terminal, option), and Range 1 and Range 2 are switched.
Remote range input terminal
Range2
Range1
Display/ AOSetting
Inst.
Wet
Gas
Avg.
HCℓ
Inst.
Wet
cv
Inst.
Dry
HC ℓ
Display/ AOSetting
Inst.
Wet
Gas
Avg.
Inst.
Wet
HCℓ
cv
Inst.
Dry
HC ℓ
2.0s or more
2.0s
2007-4-1
12:00:00
Range Control
Dry
Range1 0 ~ 50.0 ppm
Remote
Range2 0 ~ 1000ppm
Range1 0 ~ 50.0 ppm
Range2 0 ~ 1000ppm
(
)
NT
2007-4-1
12:00:00
Range Control
Dry
Range1 0 ~ 50.0 ppm
Remote
Range2 0 ~ 1000ppm
Range1 0 ~ 50.0 ppm
Range2 0 ~ 1000ppm
(
)
NT
- 56 -
2.0s or more
2.0s or more
2.0s
Select the component to be set by the
or the
key. Press the
the cursor to the "Inst./Avg." setting.
Select "Inst." or "Avg." by the
key. Press the
sor to "Wet/Dry" setting.
Short-circuit
Open
key
NT
key to move
key or the
NT
key to move the cur-
INZ-TN1ZSS-E