Fuji Electric FE ZSS Instruction Manual page 79

Cross stack laser gas analyzer
Table of Contents

Advertisement

(22) Contact output
(23) Contact input (option)
(24) Alarm output
(screen-displayed)
(25) Display contents
2. Contact output contents (1A contact)
(1) Low light transmission
(2) Outside the range of up-
per/lower limits
(3) Device failure
(4) On hold
(5) Power OFF
3. Contact input contents (option)
(1) Average value reset signal : Output of converted average value is started from the initial state by
(2) Switching instantaneous
value/moving average
value
(3) Remote hold
4. Standard functions
(1) O
correction
2
INZ-TN1ZSS-E
: Relay contact output (contact capacity 24V DC 1A 1a or 1b × 5
(Standard 1a)
Insufficient amount of light received, outside the range of up-
per/lower limits, device faulty, on hold / under calibration, power
OFF
: Photo coupler receiver contact input (operating voltage 12 to 24V DC
/ 5 to 20mA) × 3
Average value reset signal, switching instantaneous value/moving
average value and remote hold
: LD failure, LD temperature error, high gas temperature, air purge
(low pressure), air purge (high pressure), box temperature warning,
low light transmission, PD over range, connection error, AI under,
Range (H-Limit) or Range (L-Limit)
: Component, concentration (instantaneous value, average value O
correction instantaneous value and O
alarm (fault status)
: Contact output is performed (off) when the amount of light transmis-
sion is insufcient.
: According to the preset upper or lower limit alarm value, contact
output is performed (off) when it becomes lower than alarm up-
per/lower limit.
: Contact output is performed (off) when laser failure, laser tempera-
ture control failure, exceeding the amount of light transmission or
connection error are occurred.
: While AO output is held by the hold setting, the values, output (off)
of which are held, are the ones just before the holding or arbitrary set
values. Contact output is performed (off) under calibration.
: Contact output is performed (off) when power is turned off.
applying rectangular-wave voltage (pulse width 2 sec or more) to the
input terminal of the average value resetting. Output is reset by in-
putting and restarted by opening.
: Switching to and from the instantaneous value and the average value
of the analog output is performed by applying rectangular wave
voltage (pulse width 2 sec or more) to the input terminal for switch-
ing between the instantaneous value and the moving average values.
: The analog output is held by applying rectangular wave voltage
(pulse width 2 sec or more) to the remote hold input terminal, and
restarted by applying it again.
: Conversion of measured HCl gas concentrations into values at stan-
dard O
concentration
2
Correction formula:
21 – On
C =
× Cs
21 – Os
C: Converted concentration
Cs: Measured concentration of sample gas
Os: Measured O
concentration (Upper limit settable 1 to 20% O
2
On: Standard O
concentration (value changeable by setting; 0 to
2
19% O
)
2
The result of calculation is indicated and output in an analog output
signal.
- 69 -
correction average value),
2
2
)
2