1.5 Description of display and operation panels

name and description of function on the operation panel.

1.5.1 Name and description of operation panel

FUSE
250V T 3.15A
Operation panel
INZ-TN2ZSJ-E
Display unit
ESC
ENTER
1
2
MODE
4
5
7
ESC
Details of operation key
No
Name
1
MODE key
2
ZERO key
3
SPAN key
4
SIDE key
5
DOWN key
6
UP key
7
ESC key
8
ENT key
17
3
SPAN
ZERO
6
8
ENT
Description
Used to switch the mode.
Used for zero calibration.
Used for span calibration.
Used to change the selected item
(by moving the cursor) and
digit.
Used to change the selected item
(by moving the cursor) and to
decrease digit.
Used to change the selected item
(by moving the cursor) and to
increase digit.
Used to return to a previous screen
or cancel the setting midway.
Used for confirmation of selected
items or values, and for execution
of calibration.