5. Installation environment
(1) Ambient temperature
(2) Ambient humidity
(3) Measurable optical path
length (stack diameter)
(4) Standard flange
(5) Air purge
(6) Air purge ow rate
(7) Measured gas condition
(8) Dust
(9) Vibration
6. Performance
(1) Repeatability
(2) Linearity
(3) Zero drift
(4) Interference from other
gas components
(5) Minimum detectable limit
(6) Response time (90% FS
response)
7. Scope of delivery
(1) Receiver unit
(2) Transmitter unit
(3) Control unit
(4) Cable between receiver unit and transmitter unit (specified length)
(5) Cable between control unit and receiver unit (specified length)
(6) Standard accessory set, instruction manual
8. Optional items
(1)
Spare parts for one year (ZBN1SS11)
(2)
Calibration gas cell (Standard length 1m) (*1) (*2)
(3)
Receiver/transmitter cable (*1)
(4)
Cable between control unit and receiver unit cable (*1)
(5)
Standard gas (ZBM), pressure regulator (ZBD)
(6)
Recorder (when necessary, Fuji's product type PHR, etc.)
(7)
Others
*1: Although there is no need to arrange the item for each equipment, there must be at least 1 item for
each installation location.
*2: The length of calibration gas cell might be changed by measuring range.
: –20 to 55°C (Receiver unit/Transmitter unit), –5 to 45°C (control
unit)
: 90% R.H. or less
: 0.5 to 10m
: JIS10K 50A flange (JIS B 2212)
: Instrument air (compressor must be installed when power supply
cannot be provided.) Pressure 0.5 to 0.7MPa or more
: 20L/min or more (depending on the measured gas temperature, ve-
locity, pressure, moisture and dust)
: Temperature: 450°C or less
Pressure: ±10kPa
Moisture: 50vol% or less
Velocity: 10m/s or less
3
: 5 to 10g/Nm
(depending on the installation environment conditions)
: 0.5G or less (0.2G or less when the frequency range is 20 to 40Hz)
(When the optical path length is 1m.)
: ±2.0% FS
: ±3.0% FS
: ±2.0%FS / 6 months (NH
20ppm or less)
: ±2.0%FS (The gasses without absorption within the measuring wave-
length rage (several tens pm) are not interfered within principle).
: HCl 0.05ppm
NH
0.15ppm
3
: 2 seconds or less
- 70 -
±3.0%FS / 6 months when range is
3
INZ-TN1ZSS-E