Checking Received Light Quantity - Fuji Electric FE ZSS Instruction Manual

Cross stack laser gas analyzer
Table of Contents

Advertisement

(14) Adjust the light quantity so that the light quantity value at the receiver unit becomes the
neighborhood of the value recorded in (7). If the angle is adjusted with the furnace oper-
ated, it may not become the neighborhood of the light quantity value in (7) due to the
effect of dust, etc.
(15) Enter the length of the stack, referring to "6.5 Parameter setting".
(16) Set the analog output setting, referring to "6.8 Selecting analog output". Also, set other
parameters that are necessary.

4.6 Checking received light quantity

(1) Remove the probe adjusting the received light quantity (BNC receptor) of the receiver unit or the
transmitter unit, and connect it to the digital voltmeter using the digital voltage cable for connecting the
BNC receptor.
(2) Read the DC voltage value.
(3) The voltage value, where zero gas is supplied with the calibration cell connected, becomes the refer-
ence light quantity value of the equipment.
INZ-TN1ZSS-E
Fig. 4-3
- 15 -
Receiver unit
robe adjusting the
received light quantity
P.21
P.47 to 49
P.62