Fuji Electric FE ZSS Instruction Manual page 4

Cross stack laser gas analyzer
Table of Contents

Advertisement

AUTION
INZ-TN1ZSS-E
Caution on Wiring
• Be sure to connect a ground wire securely to the specified place by per-
forming class D grounding work. Otherwise electric shock or malfunction
may result.
• If the power supply voltage exceeds the rating, electric shock or damage
to the instrument may result. Be sure to use the instrument within the
specified rating range.
• Be sure to turn off the power before performing wiring work.
• Be sure to use a 600V-IV ground wire 2 mm or larger in diameter with
sufficient dielectric strength.
• Select input/output wires of materials and diameter that satisfy the rating
of each device. If a wire which cannot endure the rating is used, electric
shock or fire may occur.
• Fasten the input/output wires to the floor or wall, and use a wire protec-
tion device.
- iii -