Alarm Record" Screen; Name (Functions) - Fuji Electric FE ZSS Instruction Manual

Cross stack laser gas analyzer
Table of Contents

Advertisement

5.3.1.6 "Alarm Record" screen

Alarm Record
Date
04/01 13:23
03/29 11:01
(1)
(2)
The screen displays the alarm record occurred in the past. The ten newest errors are logged. The oldest error
will be deleted one by one every time a new alarm occurs. New errors are displayed from the top on the
screen. If the power display supply is turned OFF, the contents in the error log file will be deleted.
It displays the date, time and component when an alarm occurred, alarm contents, recovery date, and recov-
ery time from the left to right.

5.3.1.7 Name (functions)

(1) Alarm occurrence date················· Displays the date when device failure, high gas temperature or
(2) Alarm occurrence time ················ Displays the time when alarm occurred.
(3) Alarm occurrence component ····· Displays the component and external input for which alarm oc-
(4) Alarm contents ···························· Displays the contents of alarm.
(5) Alarm recovery date ···················· Displays the date when alarm is recovered. Nothing is displayed
(6) Alarm recovery time···················· Displays the time when alarm is recovered. Nothing is displayed
(7) Select Delete key ························· Deletes the selected alarm.
(8) All Delete key······························ Deletes all alarms.
Day
NH3
NH3
(3)
(7)
Select Delete
alarm occurred.
curred.
for the alarm which is not recovered.
for the alarm which is not recovered.
Alarm
Over H-Limit 1
LD Temp. Error
(4)
All Delete
- 28 -
2006-4-1
12:00:00
Day Repair
31
03:23
(5)
(8)
(6)
INZ-TN1ZSS-E