Name
(1)
Receiver box
(2)
Transmitter box
(3)
Adjustment probe for received
light quantity
(4)
Angle adjustment unit
(5)
Bolt for angle fine adjustment
(6)
Air purge inlet
(7)
Mounting flange
(8)
Receiver / Transmitter unit cable
(9)
Cable between control unit and
receiver unit
(10) Control unit fixed fixture
(11) Display part
(12) External input/output part
INZ-TN1ZSS-E
A photodiode to receive light and a board are built-in.
Laser light source, peltier to control the temperature of laser and a board
are built-in.
Terminal which checks the intensity of received light volume when light
axis adjustment is performed. It is provided in both the receiver unit and
transmitter unit. Connecting the terminal is the BNC receptor.
A unit which adjusts light volume to receive laser light transmitted from
transmitter unit via PD of receiver unit. Up to 5° can be angle-tuned by the
bolt.
When the stack becomes longer, received light quantity changes greatly by
subtle angle-turning. In such cases, light axis adjustment is performed by
using the bolt for the flange of angle adjustment or the bolt for angle fine
adjustment. It may not be necessary when the stack (optical path lengths) is
short and less than 2mm. Use it according to the length of the stack.
Inlet for feeding instrumentation air. It is required to prevent lens conden-
sation, eliminate attached dust and prevent the pipe from being clogged
with dust. Prepare the pipe to connect to joint for tube φ 8/6mm.
Connects to the companion flange which is mounted on the stack. The stan-
dard flange is JIS 10K50A.
Supplies power to transmitter unit, and sends and receives signal between
receiver unit and transmitter unit. The standard length is 2m and up to 25m
can be specified.
Supplies power to receiver unit, and sends and receives signal between
receiver unit and transmitter unit. The standard length is 5m and up to
100m can be specified.
Fixes control unit on the wall. There are two φ 12mm holes as mounting
holes at upper part, and 12mm width U-tube hole at lower part.
Displays the measurement value and alarm.
Input/output part for AC power supply, signal cable to the receiver unit,
analog input/output and relay input/output. The hole diameter is φ 10mm.