Installing The Npd Control Card; Extension Slots Top View - Thermo Scientific TRACE GC Ultra Installation Manual

Hide thumbs Also See for TRACE GC Ultra:
Table of Contents

Advertisement

Getting Started

Installing the NPD Control Card

The NPD control card must be installed into the proper expansion slot marked A,
B or C located on the left part of the GC mother board in the electronic
compartment. Refer to Figures 1-5 and 1-6.
NOTE
In the case all the expansion slots are occupied, it is necessary to remove one of the three
detector control card currently installed and replace it with the NPD control card.
Material needed:
1. Loosen the fixing screw of the slot cover plate that corresponds to the selected
Installation Guide
3-mm Allen wrench
A, B or C expansion slot on the mother board.
1. Slot cover plate
3. Free slot
C
1
B
3
A
2. Cover plates fixing screw
A B C. Extension slot position
Figure 1-5. Extension Slots Top View