7
Error Information
Pump Error Messages
Deliver underrun
Error ID: 29233
Internal error.
• Parameter: None
Probable cause
1
2
Defect connection between main board and pump drive encoder
Error ID: 29208
Defect connection between main board and pump drive encoder.
• Parameter: 1 – 2 referring to pump drive
Probable cause
1
2
Pump drive encoder defect
Error ID: 29209
Defect pump drive encoder.
• Parameter: 1 – 2 referring to pump drive
Probable cause
1
122
Internal error.
Firmware issue
Defect connection between main board and
pump drive encoder.
Defect encoder.
Defect encoder.
Suggested actions
Please contact your Agilent service
representative.
Use a minimum firmware revision of B.06.55
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Suggested actions
Please contact your Agilent service
representative.
1290 Infinity Quaternary Pump User Manual